Yd URiSCAN Optima Urine Analyzer
Configuration Guide
Step 1: Add device to the Relaymed Software Agent
To add the URiSCAN Optima (manufacturer: Yd Diagnostics) to the Relaylink follow the Add Ethernet device to Relaylink guide
Step 2: Connect device to the network
- Connect the Male-to-Male Null Modem adapter to the device and to the Lantronix xDirect
- Connect the Lantronix xDirect to the RS232 Port on the rear of the device
- Follow the Lantronix xDirect configuration guide to correctly configure the adapter to connect to your network
Step 3: Setup the device
- From the Main Menu select "Setting"
- From the parameters menu select "Set Comm."
- Set serial options to:
- Braudrate: 9600
- Data bits: 8
- Parity: None
- Stop bits: 1
Step 4: Send a test through
To verify the setup works send a test from the device to Relaymed.