Siemens Clinitek Advantus
Configuration Guide
Configuration Stages
To configure the Siemens Clinitek Advantus to connect to Relaymed the following configuration stages are required:
- Configure the Siemens Clinitek Advantus Device Settings (On the Device)
- Configure the Remote Host connection settings (Via Web Browser)
- Add device to Relaylink.
Stage One: Device Configuration
Step 1: From the Siemens Clinitek Advantus Home Screen, select "Menu":
Step 2: From the Advantus Main Menu, select "Setup":
Step 3: Scroll through the Setup Menu using the "Right Arrow" screens to find "Network settings"
Step 4: Select the "Network settings" option from the Setup Menu.
Step 5: Configure the Siemens Clinitek Advantus with the required network settings.
NOTE: The Clinitek Advantus's network settings are split over two pages, use the Right Arrow to navigate to the second page.
Step 6: Save the Network settings and return to the Setup menu by selecting the "Left Arrow" to return to the first Network settings page and then select the "Left Arrow" again to return to the Setup menu.
Step 7: Select "Computer port options" option from the Setup menu.
Step 8: Ensure the Siemens Clinitek Advantus Computer port options are configured to the following settings:
- Port - ETHERNET PORT
- Baud - 9600
- Data, Parity - 8/NONE
- Output Format: CSS
Step 9: Once the options are correctly configured select the "Left Arrow" to save the settings and return to the Setup menu.
Step 10: Select the "Home" button to return to the Clinitek Advantus Home screen. This complete Stage One of the Clinitek Advantus configuration.
Stage Two - Remote Host Configuration
Step 1: On a Computer connected to the same network as the Clinitek Advantus launch your web-browser (Google Chrome Recommended). In the Address bar enter the IP Address allocated to the Siemens Clinitek Advantus to access the configuration portal
Step 2: When prompted enter the User Name and Password to access the Clinitek Advantus configuration portal.
Step 3: In the configuration portal select "Connection" from the Menu.
Step 4: Ensure the Siemens Clinitek Advantus Connection options are configured to the following settings:
- Connect Mode>Passive Connection>Accept incoming : Yes
- Connect Mode>Active Connection>Active: Auto Start
- Connect Mode>Password Required: No
- Endpoint Configuration>Remote Port: Enter the Port allocated to the Clinitek Advantus in the Relaylink (local port should remain default).
- Endpoint Configuration>Remote Host: Enter the Static IP Address of the Host Machine with the Relaylink installed
- Disconnect Mode>Hard Disconnect : No
Step 5: Click "OK" to save the Connection settings.
Step 6: Click "Apply settings" to apply the settings to the Clinitek Advantus
Step 7: The Clinitek Advantus's Network Controller will restart to apply the settings.
Stage Three: Add Device to Relaylink
To add the Siemens Clinitek Advantus to the Relaylink follow the Add Ethernet device to Relaylink guide