Quidel Triage MeterPro

Configuration Guide

Step 0: Before you start

To Enable the LIS output on the Triage the user requires a LIS password unique to the device. This is obtained from Quidel by providing the Triage’s serial number. Make sure the Supervisor Chip is plugged in.

Step 1: Add device to the Relaylink

To add the Quidel Triage MeterPro to the Relaylink follow the Add Ethernet device to Relaylink guide

Step 2: Connect device to the network

  • Connect the Male-to-Male Null Modem adapter to the device and to the Lantronix xDirect
  • Connect the Lantronix xDirect to the RS232 Port on the rear of the device
  • Follow the Lantronix xDirect configuration guide to correctly configure the adapter to connect to your network

Step 3: Setup the device

Enable LIS output

  • From the Main Menu select "Set parameters"
  • From the parameters menu select "Meter Settings"
  • From Meter Settings choose "Comm Settings"
  • Select "LIS", then use the Left/Right Arrows to toggle Enabled, then enter to confirm the selection
  • Enter the password provided by the Quidel™ Representative

LIS Settings

  • From the Main Menu select "Set parameters"
  • From the parameters menu select "Meter Settings"
  • From Meter Settings choose "Comm Settings".
  • Select "Baud", then use the Left/Right Arrows to display 9600
  • Select "Auto Upload", then use the Left/Right Arrows to toggle Enabled, then enter to confirm the selection

Step 4: Send a test through

To verify the setup works send a test from the device to Relaymed.