Quidel Triage MeterPro
Configuration Guide
Step 0: Before you start
To Enable the LIS output on the Triage the user requires a LIS password unique to the device. This is obtained from Quidel by providing the Triage’s serial number. Make sure the Supervisor Chip is plugged in.
Step 1: Add device to the Relaylink
To add the Quidel Triage MeterPro to the Relaylink follow the Add Ethernet device to Relaylink guide
Step 2: Connect device to the network
- Connect the Male-to-Male Null Modem adapter to the device and to the Lantronix xDirect
- Connect the Lantronix xDirect to the RS232 Port on the rear of the device
- Follow the Lantronix xDirect configuration guide to correctly configure the adapter to connect to your network
Step 3: Setup the device
Enable LIS output
- From the Main Menu select "Set parameters"
- From the parameters menu select "Meter Settings"
- From Meter Settings choose "Comm Settings"
- Select "LIS", then use the Left/Right Arrows to toggle Enabled, then enter to confirm the selection
- Enter the password provided by the Quidel™ Representative
LIS Settings
- From the Main Menu select "Set parameters"
- From the parameters menu select "Meter Settings"
- From Meter Settings choose "Comm Settings".
- Select "Baud", then use the Left/Right Arrows to display 9600
- Select "Auto Upload", then use the Left/Right Arrows to toggle Enabled, then enter to confirm the selection
Step 4: Send a test through
To verify the setup works send a test from the device to Relaymed.