Qualigen FastPack IP Version 2
Configuration Guide
Configuration Stages
To configure the Qualigen FastPack IP System to connect to Relaymed the following configuration stages are required:
- Configure the Qualigen FastPack IP System Device Settings
- Qualigen FastPack IP System Connection Configuration
- Lantronix Configuration for Ethernet
- Digi Edgeport connection for Direct PC
- Add device to Relaylink.
Stage One: Configure Device
To allow the Qualigen FastPack IP System to communicate with Relaymed the following settings are required:
- Language Units: English 2
Step 1: From the Main Menu use the Red button to cycle through menu options to Language Units setting.
Step 2: Press the Blue button to enter the Language Units menu.
Step 3: Press the Red button to cycle through the options until "English 2" is displayed.
Step 4: Press the Blue button to save the setting.
Step 5: Press the Red button until "Exit" is displayed.
Step 6: Press the Blue button to exit the menu and save the settings.
Ethernet Connection
Stage Two: Lantronix Configuration (Ethernet Connection Only)
Step 1: Connect the Male-to-Male gender changer adapter to the COM 2 port on the rear of the Qualigen FastPack IP.
Step 2: Connect the Lantronix xDirect to the Male-to-Male gender changer adapter.
Step 3: Follow the Lantronix xDirect configuration guide to correctly configure the adapter to connect to your network.
Stage Three: Add Device to Relaylink
To add the Qualigen FastPack IP to the Relaylink follow the Add Ethernet device to Relaylink guide
Direct PC Connection
Stage Two: Digi Edgeport Connection (Direct PC Connection Only)
Step 1: Connect the Digi Edgeport to to the COM 2 port on the rear of the Qualigen FastPack IP. and an available USB port on the PC.
Step 2: Record the Serial Number on the Digi Edgeport Adapter, this will be required to add the device to the Relaylink.
Stage Three: Add Device to Relaylink
To add the Qualigen FastPack IP to the Relaylink follow the Add Serial device to Relaylink guide