Horiba Pentra 60 C+
Configuration Guide
To configure the Horiba Pentra 60 C+ to connect to Relaymed the following configuration stages are required:
- Horiba Pentra C+ 60 Connection Configuration
- Lantronix xDirect Adapter Configuration
- Add device to Relaylink.
Stage One: Connection Configuration*
*This guide is still being developed, please amend any information that may not be accurate as new info becomes available
Step 1: From the connected computer navigate to the Settings menu and then System Setting
Step 2: Select the Communication tab and ensure the following settings are selected:
RS232 Configuration
- Bauds: 9600
- Parity: None
- Stop Bit: 1
Sending Configuration
- Format: ASTM
- Mode: UNIDIR
Step 3: Send a test over using the recall functionality of the device
Stage Two: Lantronix Configuration
Step 1: Connect the Lantronix xDirect directly to the RS232 Port.
Step 2: Follow the Lantronix xDirect configuration guide to correctly configure the adapter to connect to your network.
Stage Three: Add Device to Relaylink
To add the Horiba ABX Micros 60 to the Relaylink follow the Add Ethernet device to Relaylink guide