CoaguSense, Inc. Coag-Sense PT2
Configuration Guide
Step 1: Add Device to the Relaymed Software Agent
Minimum Relaylink version required 3.3.0
Further assistance on Step One can be found here: Add Ethernet device to Relaylink
Step 2: Configure Device Network Settings
1. Make sure the user is logged in under the administrator account. The administrator login info is provided by the manufacturer at the time of purchase and is included in the package when shipped.
2. Select the Settings menu and then select Data Transfer. Select POCT1-A, select Confirm to save and return to the previous menu.
3. Ensure that the ethernet cable is plugged in to the CAT-45 Port on the back of the device to use this 'Wired' selection.
4. Look for Communication settings and select Wired Connection.
5. Under Wired settings select either DHCP or Manual (Static). If DHCP is selected no further action is needed. If Manual (Static) IP is selected you will need to fill in the required fields in the menu. This information can be found by contacting your IT Dept or third-party IT Support. Select Confirm to save and return.
6. Select Server and input the Host IP address, this is the computer/server on which the Relaymed software agent is installed. Next input the Port number that was entered when setting up the device on the Relaymed software agent. This information can be found by contacting your IT Dept or third-party IT Support. Once completed select Connect.
Note: A restart of the device is suggested and may be required.
Step 3: Send a test
Verify setup is working by sending a test to Relaymed.
Test data is sent automatically once the device completes the test.