Alfa Wassermann ACE Axcel 

Configuration Guide

Configuration Stages

To configure the Alfa Wassermann Ace Axcel Clinical Chemistry System to connect to Relaymed, complete the following stages:

  1. Ace Axcel System Connection Configuration
  2. Serial to Ethernet Adapter Configuration
  3. Add Device to Relaymed

Stage 1: Connection Configuration

To enable communication between the Ace Axcel and Relaymed, follow these steps:

Step 1:
From the Main Menu, select “Setup.”

Step 2:
Within Setup, select “Hardware.”

Step 3:
On the Hardware screen, under the LIMS section, ensure the following settings are configured:

  • Enable: Yes
  • Baud Rate: 9600
  • Parity: None
  • Data Bits: 8
  • Stop Bits: 1

Step 4:
Select “Accept (F10)” to save your changes.


Stage 2: Serial to Ethernet  Configuration

Step 1:
Connect the Serial to Ethernet adapter directly to the RS232 port labeled “LIMS” on the back of the Ace Axcel device.

Step 2:
Follow the Serial to Ethernet configuration guide to properly set up the adapter and connect it to your network.


Stage 3: Add Device to Relaymed

For practices using the Relaymed Link, log in to your Relaymed Portal and follow these steps:

Adding a Device to Relaymed Link

For practices using the Relaymed Software Agent, open the server or PC where the agent is installed and follow these steps:

Adding a Device to Software Agent