Alfa Wassermann ACE Axcel 

Configuration Guide

Device Summary Info

Configuration Stages

To configure the Alfa Wassermann Ace Axcel Clinical Chemistry System to connect to Relaymed the following configuration stages are required:

  1. Ace Axcel Clinical Chemistry System Connection Configuration
  2. Lantronix xDirect Adapter Configuration
  3. Add device to Relaylink.

Stage One: Connection Configuration

To allow the Alfa Wassermann Ace Axcel to communicate with Relaymed the following steps are required.

Step 1: From the  "Main Menu" select the "Setup" option.

Step 2:  From the "Setup " select "Hardware".

Step 3: On the "Hardware" screen under the "LIMS" section ensure the following options are selected:

  • ENABLE: YES
  • BAUD RATE: 9600
  • PARITY: NONE
  • DATA BITS: 8
  • STOP BITS: 1

AlfaHArdwareMenu.jpg

Step 4:  One the "Hardware" screen select "Accept (F10)" to save the changes.

Stage Two: Lantronix Configuration

Step 1: Connect the Lantronix xDirect to the directly to the RS232 Port labelled "LIMS" on the rear of the Alfa Wassermann Ace Axcel.

Step 2: Follow the Lantronix xDirect configuration guide to correctly configure the adapter to connect to your network.

Stage Three: Add Device to Relaylink

To add the Alfa Wassermann Ace Axcel to the Relaylink follow the Add Ethernet device to Relaylink guide